Home
Finger Blood Pressure Monitor
Instruction Booklet
Read All Instructions Carefully Before Use
DO’S AND DON’TS
FOR ACCURATE BLOOD PRESSURE READINGS
1. DO Read instructions carefully before using Monitor.
2. DO Rest arm on a flat surface so that finger is at heart level
when taking measurements.
3. DO Remain calm, quiet and still when measurements are being
taken.
4. DO Make sure that the Finger Cuff is adjusted properly according
to your instruction manual.
5. DO Insure that the exhaust velocity is 2-5 mmHg per second at
the systolic point each time a measurement is taken.
6. DO Wait at least 10 minutes between taking measurements.
7. DO Pump unit 30 to 60 mmHg above normal systolic before beginning measurement.
8. DO Consult your physician before adjusting your medication based on readings from
this Monitor.
9. DO Use fresh alkaline batteries.
IO. DO Use a dry, soft cloth to clean the instrument.
11. DO use your left index finger when taking measurement. Use of an alternate finger may
result in an inaccurate reading.
12. DO make sure that your finger is warm.
1. DO NOT move arms or hands when measurement is being taken.
2. DO NOT talk or chew gum when measurement is being taken.
3. DO NOT take a series of measurements without waiting at least 10 minutes
between
measurements.
4. DO NOT adjust medication based on readings from this Monitor.
5. DO NOT make any adjustments to unit; there are no user serviceable parts. Return to
Service Station for repair.
6. DO NOT smoke during or at least 30 minutes prior to taking blood pressure measurements.
7. DO NOT drop or otherwise mechanically shock the Monitor. This is a delicate
precision device.
8. DO NOT use thinner, alcohol, benzene or wet cloth to clean unit.
9. DO NOT subject to extremes in temperature, humidity, direct sunlight or dust.
Manufactured to meet specifications set by the Association for the Advancement of
Medical Instrumentation (AAMI)
WHAT IS BLOOD PRESSURE?
Measurement of the rate of pressure exerted by the blood upon the
walls of the blood vessels. The heart contracts and then
relaxes, producing the two levels of blood pressure.
Bloodpressure:
The highest point of blood pressure; level of blood pressure
produced as your heart beats.
Systolic:
Diastolic:
The lowest point of blood pressure; level of blood pressure
produced as your heart rests between beats.
Millimeters of Mercury (mmHg):
Blood pressure measurements are measured in terms of the
height of mercury in a column. Measurements are written one
over the other. For example, a systolic pressure of 120 mmHg and
diastolic pressure of 80 mmHg is expressed as “120 over 80.”
Blood pressure varies constantly and may change due to:
l
l
l
-----
l
l
physical trauma
mental anxieties
smoking
exercise
eating
drinking caffeinated or alcoholic beverages
Only a qualified physician can determine whether your blood pressure is normal.
is high blood pressure, and is most common among adults and the
elderly; left unattended, it can cause serious health problems such as a
stroke, heart attack, etc. Therefore, it is important to monitor blood
pressure on a continued and timely basis.
Hypertension:
Only a physician is qualified to interpret your blood pressure
measurements, and no device can replace regular medical exami-
nations by your physician. It is recommended that your physician
review your procedure for using this blood pressure monitor. Your
physician should verify blood pressure measurements before
DomwnalokadinfrgomaWdwjwu.sSotmmaenunatlss.cotmo. Amll eMdaniucaalstSioeanrc.h And Download.
IMPORTANT:
BLOOD PRESSURE RANGES
In young, healthy adults the normal blood pressure range is between 100 and 140 mmHg (milli-
meters of mercury) for SYSTOLIC pressure, and between 60 and 90 mmHg for DIASTOLIC
pressure.
The range of pressure can vary considerably depending on age and physical condition.
ONLY A QUALIFIED PHYSICIAN CAN DETERMINE WHAT IS “NORMAL” FOR A
PARTICULAR
PERSON.
activity
Your reading can also vary during the .day and depending on your physical condition and
level. In general, readings will be lower when you are calm and rested, and higher following
physical activity. There will be some variance in consecutive readings or readings from one arm
to the other.
For these reasons, it is recommended that blood pressure be measured after sitting or
resting quietly for 30 minutes or more following any strenuous activity.
To minimize variations, readings should be taken at similar times and under similar conditions of
stress. Record daily readings for accurate comparisons.
PARTS OF YOUR BLOOD PRESSURE MONITOR
Cuff Size
Indicator
Finger Cuff
\
Start
Button
On/Off
Button
Blood Pressure and
Pulse Rate Display
DESCRIPTION OF DISPLAY MARKS
CORRECTIVE ACTION
CONDITION/CAUSE
,
DISPLAY MARK
Measurement in progress; re-
main quiet.
Heart mark appears during
measurement conditions and
flashes when pulse is de-
tected.
Measurement in progress
Automatic exhaust occurs.
Mark flashes when power is
turned on and there is air re-
maining in the Cuff. Mark
flashes at the completion of
the measurement until ex-
haust is complete.
Exhausting
Set the ON/OFF Button to Off.
Set it back to ON, and then
start the measurement again
when “0” appears.
Measurement began, but the
pressure was insufficient.
Insufficient Pressure
Appears when the battery is
excessively low.
Replace all four batteries with
new ones. (4
-
AAA
Replace Batteries
Alkaline batteries)
Appears when the blood
pressure cannot be mea-
sured accurately.
Press ON/OFF Button to turn
OFF. Press again to ON and
start the measurement again
when “0” appears.
Measurement
Error
The pulse is very weak; or
the finger is cold; or Finger
Cuff is too loose.
If you have poor circulation in
your hand, you will not get a
reading. Warm up your hand
and retry measurement.
Tighten Finger Cuff and retry
measurement.
E
r
r
BATTERY INSTALLATION
4 AAA alkaline batteries are required. Remove the
battery cover and insert batteries into the battery
. compartment on back of monitor. (Fig.1) Be sure that
you insert the batteries according to the polarities
specified on the bottom of the battery compartment. If
Monitor is to be stored for extended periods, remove the
batteries.
Figure 1
HOW TO USE YOUR SUNBEAM DIGITAL
BLOOD PRESSURE MONITOR
1. Slide the Cuff Size Adjuster to the far left (L position)
to open the finger cuff fully.
2. Insert your left index finger fully into the Finger Cuff.
(Use of an alternate finger will result in an improper
reading) (Fig.2) Be sure your finger is warm and
relaxed. (If your finger is cold an improper reading may
result)
Figure
2
3. Press the ON/OFF Button. Three sets of symbols will
appear in rapid sequence on the display. (Fig. 3) Wait
for the "O" symbol to appear in the display before
starting measurement, indicating the unit is ready to
measure.
4. Slide the Cuff Size Adjuster to the right until the Cuff
Size Indicator light turns on. When the red light turns
on, this means that the Finger Cuff is too tight. If your
fingers are slender, the smallest finger cuff position may
be the most suitable. In this case, the red Too Tight
Indicator will not light up. An inaccurate blood pressure
reading may result from improper Finger Cuff adjust-
ment.
5. Hold the Monitor steady at heart level with one or
both hands. Be sure that when holding the Monitor, your
left hand is relaxed. Press the Start Button. The Finger
Cuff will automatically inflate to a pressure of about 200
mmHg. If a higher starting point than 2OOmmHg is
required, the monitor will stop inflation automatically,
then resume inflation after determining the correct
Figure 3
starting point.
6. Remain still after inflation is completed. The Monitor
will then begin deflation of the Finger Bladder automati-
cally. Pressure readings will be indicated in the display
as the pressure in the Finger Bladder decreases. As the
Monitor continues to deflate the Finger Bladder,
a
heart
will start to flash in the display. (Fig. 4)
7. The Monitor will deflate automatically when measure-
ment is complete, and the resulting blood pressure
readings will be indicated in the display. Blood pressure
readings (Systolic on the left, Diastolic on the right), will
be displayed alternately with the pulse (Pul) reading.
(Fig. 5)
Figure 4
8. Display of measurement results will continue until the
power is turned off. The Monitor will shut off automati-
cally after approximately
2
minutes.
9. Wait at least 10 minutes between readings. To
discontinue measurement for any reason, press the
a
ON/OFF Button. This causes the finger cuff to deflate
rapidly. Before removing your finger from the Finger
Cuff, slide the Cuff Adjuster to the left. Failure to do
Figure 5
so could result in damage to the Finger Cuff.
SPECIFICATIONS
Model: 7655-10
Power Source: Type AAA batteries (4)
Battery Life: Approx. 6 months with 3 min. usage per day
Power Consumption: 600mW
Type: Oscillometric
Measurement Range:
0
-
300mmHg (Pressure)
Operating Environment:
-
-
40
-
150 pulses/minute (Pulse)
30
Storage Environment: -20°C
30
-
85%RH
Accuracy: +I-3mmHg or +/-2% of reading whichever
-
is greater
-
95% RH
x
Dimensions: 4”(W) x
Pulse:
+/- 5% of reading
Finger Circumference Range: Approx. 2-l/8” to 3”
Weight w/batteries: Approx. 240gm.
Origin: Made in Japan
Inflation: Automatic, using an electronic pump
Deflation: Automatic pressure release valve
If you have any questions about your new blood pressure monitor, please call our toll free number for
assistance at l-800-621 -8854.
LIMITED ONE YEAR WARRANTY
This product is warranted for one year from the date of purchase against defects in material and workmanship. If the product
fails to operate during this period, return it, prepaid, to Service Center 117 Central Industrial Row Purvis, MS 39475
for repair or replacement without charge at the manufacturer’s option. This warranty gives you specific legal
rights. and you may also have other rights which vary from state to state. Warrantor: Sunbeam Home Comfort
@Sunbeam TM Health at Home, 01992 Sunbeam Home Comfort Distributed by Sunbeam Home Comfort, Schaumburg, IL 60173
P.N. 47834-025-000
|