Roper Home Theater Server llc 200 User Guide

4411-0068  
Version 1.A  
December 10, 1997  
ꢀꢁꢁꢂꢂꢃꢄꢄꢅꢆꢀ  
 
Table of Contents  
Model IIC-200 Image Intensifier Controller........................................................ 5  
Introduction.........................................................................................................................5  
Dangers and Warnings........................................................................................................6  
Features...............................................................................................................................7  
Specifications......................................................................................................................7  
Front Panel Features ...........................................................................................................8  
Back Panel Features............................................................................................................8  
Line Voltage Selection........................................................................................................9  
Hardware Overview..........................................................................................................10  
Operation...........................................................................................................................11  
Connections with a Gen II Intensifier ........................................................................11  
Gated Operation...................................................................................................11  
Shutter Mode Operation.......................................................................................11  
Connections with a Gen IV Intensifier.......................................................................12  
Gated Operation...................................................................................................12  
CW or Shutter Operation .....................................................................................12  
Power Up Steps ..........................................................................................................13  
The High Voltage Cable ...................................................................................................13  
Warranty & Service............................................................................................ 15  
Warranty............................................................................................................................15  
Equipment Repairs............................................................................................................15  
Contact Information..........................................................................................................16  
Index ................................................................................................................... 17  
iii  
 
iv  
Model IIC-200 Operation Manual  
Version 1.A  
This page intentionally left blank.  
 
Model IIC-200  
Image Intensifier Controller  
Figure 1.  
Model IIC-200  
Image  
Intensifier  
Controller  
Introduction  
The IIC-200 is a stand-alone image intensifier controller. It supplies high-voltage DC  
bias for Princeton Instruments image intensifiers plus circuitry to facilitate gating.  
Gating pulses are provided by a Princeton Instruments high voltage pulser.  
The IIC-200 allows front-panel control of the MCP gain (MCP voltage). In applications  
where the average light level varies significantly, the automatic brightness control mode  
can be switched on. In this mode, the MCP gain varies inversely with the signal level,  
enhancing the input dynamic range while restricting the output dynamic range, and  
facilitating focusing. An alarm feature automatically shuts down intensifier operation in  
case of excessive phosphor current.  
5
 
6
Model IIC-200 Operation Manual  
Version 1.A  
The IIC-200 allows both Shutter and Gated modes of operation. Gating pulses can range  
from 5 nsec to DC. Longer gate experiments can also use the 5 volt logic electronic  
shutter control.  
Dangers and Warnings  
DANGERS  
Voltages inside the IIC-200 may exceed 10,000 volts. To avoid possible hazard to  
personnel, use the instrument only with the proper connectors and cables and only for the  
purposes for which it is designed, i.e. the operation of image intensifier tubes which are  
contained in proper housings and equipped with compatible connectors. Adequate  
precautions must be taken to prevent persons from making contact with the output of the  
IIC-200.  
Never plug the AC cord into an outlet which is not properly grounded. Never attempt to  
operate the unit without its covers in place.  
WARNINGS  
Never operate the unit with the high voltage cable disconnected. The high voltage  
connectors are designed to seal out air to avoid arcing. An open high voltage connector  
is a danger to personnel and can cause damage to the equipment. Connections should be  
made before power is turned on.  
Before disconnecting any cables, turn the power off and wait at least 5 minutes for  
voltages to decay to safe levels.  
The IIC-200 chassis uses a three wire power cord for connection to the power source and  
to a protective ground. The exposed parts of the instrument, including the BNC shields,  
are connected to the outlet ground to protect against electrical shock. Always use an  
outlet which has a properly connected protective ground. Never make connections to  
other instruments which might have voltages on BNC shields. Disconnecting the  
protective earth terminal, inside or outside the chassis, or tampering with its operation  
may render the IIC-200 dangerous and is prohibited.  
The high-voltage Gate In connector (BNC with Gen II intensifier, SHV with Gen IV  
intensifier) is located next to the Shutter In BNC connector on the back panel. The  
Shutter In BNC connector is the input for a TTL (5 V) logic level signal and should  
never be connected to a high-voltage gate pulse or damage could occur. Always take  
care not to confuse these inputs.  
During manufacture, a Model IIC-200 is configured for operation with a Gen II  
Intensifier or with a Gen IV Intensifier, but not for operation with both types. The two  
types are not interchangeable. Units configured for operation with a Gen IV Intensifier  
will ordinarily have a green identification label on the back panel. Units configured for  
operation with a Gen II Intensifier will have a blue label. Your detector may have  
corresponding color-coded labeling. Check with the factory if you are in doubt. Using the  
wrong IIC-200 type with your intensifier could damage the intensifier. Damage due to  
such misuse is not covered by the Warranty.  
 
Version 1.A  
Model IIC-200 Operation Manual  
7
Note: The Model IIC-200 has automatic circuitry to help prevent damage. However, it  
is far better not to place an expensive intensifier at risk than to depend on the protection  
circuitry.  
Features  
Independently variable MCP power supply, independent of phosphor voltage.  
Automatic alarm and shutdown in case of excessive phosphor current.  
Automatic brightness control switch for ease of focus.  
Gate input for gating between 5 nsec and DC, electronic shuttering input for long  
gate times or CW operation.  
Compatibility with Princeton Instruments gate pulsers and controllers.  
Specifications  
MCP power supply: 500 to 1000 volts, 100 µA maximum. Contact the factory for  
higher current applications up to 500 µA.  
Intensifier bias levels (Standard Configuration):  
Phosphor: 0 volts, i.e. grounded phosphor.  
MCP-Out: -5500 volts typical, Gen II  
MCP-In: -6000 to -6500 volts typical, Gen II  
Photocathode to MCP-In voltage: -180 volts in CW mode, +40 volts in gate  
mode, Gen II. Gen IV versions have voltages appropriate to Gen IV intensifiers.  
Gate pulse  
Gen II Intensifier: -180 to -220 V, 5 nsec to 1 msec, externally supplied.  
Princeton Instruments PG-200 Pulser recommended.  
Gen IV Intensifier: Approximately -1000 V, 12 nsec to 1 msec, externally  
supplied. BNC Corp. Model 6040 Pulse Generator with Model 310H plug-in  
module recommended (12 ns to 10 µs).  
Shutter pulse: 5 volt TTL input. Logic 1 turns the Intensifier on in CW mode.  
Logic 0 (0 V) turns it off. Used to achieve gate times longer than the 80 ms delay +  
gate width obtainable with a PG-200. Minimum TTL 1 pulse width to turn intensifier  
on is 50 sec.  
Automatic Brightness Control: Gain reduction of 16 typical under maximum  
incident light.  
Phosphor over-current shutdown: 30 nA typical. Over-current condition  
sounds audio beeper.  
Audio beeper: ~1 kHz, cycles on/off at rate dependent on overload.  
 
8
Model IIC-200 Operation Manual  
Version 1.A  
Front Panel Features  
MCP GAIN: This adjustment is used to control the gain of the MCP by controlling the  
voltage across it. Display units are arbitrary.  
SHUTTER/GATE: This switch selects between Gate mode or Shutter mode. In Gate  
mode, the intensifier is normally off and a high-voltage gate pulse is needed to  
turn it on. In the Shutter mode, the intensifier is normally on, and a TTL logic  
low (0 V) must be applied to the back panel Shutter In BNC connector to turn  
the intensifier off. The Shutter Monitor output provided by Princeton  
Instruments CCD Controllers can be applied to turn off the intensifier during  
readout.  
AUTO BRIGHT CONTROL ON/OFF: This switch turns the automatic brightness  
control feature on or off. Brightness control reduces the MCP gain when  
exposed to high light levels. This compresses the output dynamic range to keep  
signal levels on screen during set-up.  
Note: When making quantitative measurements, AUTO BRIGHT should be  
OFF.  
MCP POWER ON/OFF: This switch turns the intensifier voltages on or off while  
chassis power remains on.  
POWER ON/OFF: This switch turns the IIC-200 chassis power on or off.  
Back Panel Features  
GATE IN: This is the input for high voltage (-200 V) gate pulses used to turn on the  
intensifier. In units intended for use with a Gen II intensifier, GATE IN is a  
BNC connector. In units supplied for operation with a Gen IV Intensifier, which  
requires a gate pulse of approximately -1000 V, GATE IN is a type SHV  
connector. A high-voltage pulser is required in both cases.  
SHUTTER IN BNC: This TTL (5 V) input is used to synchronize the intensifier with  
the CCD exposure time or to hold the intensifier on in Shutter mode operation.  
WARNING  
If the front-panel Shutter/Gate switch is set to SHUTTER, the intensifier will be biased  
on if the Shutter In connector is floating, or if +5 V is applied. As a result, the intensifier  
is particularly vulnerable to overload damage in this mode. If the switch is set to GATE,  
the Shutter In connector is inactive and the intensifier can only be biased on for the  
duration of high-voltage gate pulses applied to the Gate In connector.  
INTENSIFIER H.V.P.S: This connector is used to supply High Voltage DC power to  
the intensifier through the cable supplied by Roper Scientific. The high voltage  
cable must always be connected at both ends before turning the power on.  
POWER MODULE: This module contains the line-cord socket, line fuse and line  
voltage selection assembly. The power and fuse requirements are printed on the  
panel above the power module.  
 
Version 1.A  
Model IIC-200 Operation Manual  
9
Figure 2.  
Model IIC-200  
back panel  
The plug on the line cord supplied with the system should be compatible with the  
line-voltage outlets in common use in the region to which the system is shipped. If  
the line cord plug should prove to be incompatible, a compatible plug should be  
installed, taking care to maintain the proper polarity to protect the equipment and  
assure user safety.  
There is provision for changing the selected line voltage and for changing the fuse.  
The line fuse value depends on the line voltage as indicated above the Power  
Module. Figure 3 illustrates the fuse access and voltage selection mechanisms. The  
Power Module is designed to accept one ¼ × 1-1/8 fuse or two 5 × 20 mm fuses.  
Line Voltage Selection  
If the IIC-200 has been in operation, turn it off and wait at least 5 minutes for  
voltages to decay to safe levels.  
Disconnect the power cord.  
Insert a small screwdriver under the cover tab at the top of the Power Module.  
Gently pry open the voltage selector cover and rotate the voltage selector until the  
setting for the local line voltage is facing you.  
Check that the appropriate fuse(s) have been installed. (1/2 Ampere T time delay for  
100/120 V AC; ¼ Ampere, T time delay for 220/240 V AC).  
Close the cover making sure the desired voltage setting appears in the window.  
Normally, the correct voltage setting and line fuses for your area will be installed at  
the factory, provided that information was specified in the order.  
 
10  
Model IIC-200 Operation Manual  
Version 1.A  
Figure 3.  
Power Module  
line voltage  
selection and  
fuse installation  
Hardware Overview  
The IIC-200 consists of a printed circuit board and a modular high voltage supply which  
are mounted in a stand-alone chassis.  
The IIC-200 contains circuitry for the following functions:  
A variable high voltage power supply for the MCP voltage. The MCP voltage is a  
linear function of the voltage supplied by the MCP Gain potentiometer.  
Upper and lower limit circuits for the MCP voltage (normally set to -500 V and  
-800 V respectively for Gen II).  
Note: When the image intensifier and IIC-200 are purchased from Roper Scientific  
as part of the same order, the upper limit circuit will be set for the maximum limit of  
the intensifier. The upper limit circuit also provides a slow turn-on during power-up.  
High-voltage adjustment circuit for the voltage between MCP-Out and the Phosphor  
(normally set for -5500 V for Gen II Intensifiers).  
Alarm and shutdown circuit to shut down the high voltage and bring the MCP  
voltage to almost zero if an over-current condition exists due to excessive light on  
the Photocathode. This circuit additionally actuates an audible alarm if an over-  
current condition occurs. The circuit automatically resets and retries after about 0.5  
seconds. Thus it will not prevent intensifier damage if excessive light is allowed to  
continuously fall on the intensifier.  
Automatic Brightness Control circuit to reduce the MCP voltage in response to  
increasing light intensity on the Photocathode. This can be useful in focusing. When  
properly interfaced to a Princeton Instruments CCD Controller, the Brightness  
Control circuit uses a sample/hold to sample the phosphor current during CCD  
exposure and hold it between exposures. The sample/hold is inactive in the Gate  
mode. The automatic Brightness Control function should be turned off for normal  
operation, since the response isn’t linear (and may not even be constant) when it is  
used.  
 
Version 1.A  
Model IIC-200 Operation Manual  
11  
Shutter circuitry which allows an electrically isolated logic signal to control the  
voltage between the Photocathode and MCP In. This allows long gate times or CW  
operation.  
Operation  
Review the Danger and Warning notices on page 6 before operating the IIC-200.  
Set the IIC-200 Power switch to OFF and wait 5 minutes.  
The high voltage cable should be handled with care. Dropping the cable, banging the  
connectors, twisting the wires or applying excessive force when tightening may damage  
the pins, resulting in a poor or intermittent connection.  
CAUTION  
Connections with a Gen II Intensifier  
Connect one end of the high-voltage cable supplied by Roper Scientific to the high-  
voltage connector on the image intensifier. If you have purchased more than one  
image intensifier, make sure you are using the cable supplied with that particular  
intensifier. The connectors are keyed and must be properly oriented. Push the cable  
connector into the receptacle and continue to turn the threaded ring clockwise until it  
is tight (10 inch-lbs maximum torque).  
Note: Two versions of the 6-pin high voltage cable have been shipped by Roper  
Scientific. Early high-voltage cables have a white housing. Later ones have a black  
housing and a white stripe. These cables are wired differently. The IIC-200 is  
compatible with the newer style cables only. For more information, see The High  
Voltage Cable on page 13.  
Connect the opposite end of the cable to the high-voltage output connector of the  
IIC-200 using the same procedure.  
Gated Operation  
For gating, set the Shutter/Gate switch to Gate.  
Connect a cable from the output of the pulser to the Gate In BNC connector. For  
Gen II units, the mating connector must be a BNC type.  
Shutter Mode Operation  
For shuttering, set the Shutter/Gate switch to Shutter.  
Connect a BNC cable to the IIC-200 Shutter In BNC connector. Connect the other  
end to the Controller’s TTL SHUTTER OUT BNC connector (may also be named  
SHUTTER or SHUTTER MONITOR, depending on the controller model) .  
WARNING  
Unlike Princeton Instruments integrated ICCD detectors, in Shutter mode operation, if  
the Shutter In BNC on the IIC-200 is allowed to float, the intensifier will be On  
Continuously. Overexposing the image intensifier at any time when the intensifier is  
turned on can result in permanent damage to the image intensifier.  
 
12  
Model IIC-200 Operation Manual  
Version 1.A  
Connections with a Gen IV Intensifier  
Connect one end of the high-voltage cable supplied by Roper Scientific to the  
high-voltage connector on the image intensifier. If you have purchased more than  
one image intensifier, make sure you are using the cable supplied with this  
intensifier. The connectors are keyed and must be properly oriented. Push the cable  
connector into the receptacle and continue to turn the threaded ring clockwise until it  
is tight (10 inch-lbs maximum torque). Connect the opposite end of the cable to the  
high-voltage output connector of the IIC-200 using the same procedure.  
Note: Two versions of the 6-pin high voltage cable have been shipped by Roper  
Scientific. Early high-voltage cables have a white housing. Later ones have a black  
housing and a white stripe. These cables are wired differently. The IIC-200 is  
compatible with the newer style cables only. For more information, see The High  
Voltage Cable on page 13.  
Gated Operation  
For gating, set the Shutter/Gate switch to Gate.  
Connect a SHV Tee (A SHV connector in the shape of the letter “T”) to the to the  
Gate In SHV connector of the IIC-200.  
Be sure the Pulser power is OFF.  
Connect a cable from one side of the SHV Tee to the HV output of a Berkeley  
Nucleonics Corp. Model 6040 Universal Pulse Generator equipped with a Model 310  
plug-in Electrical Module. This cable should be no more than 1 m long. A type N-to-  
SHV adapter will be required to make the connection. This adapter is provided if the  
6040 is ordered from Roper Scientific.  
Note: Although SHV and BNC connectors are superficially similar, they are  
dimensionally different and incompatible. The SHV connector is designed for high-  
voltage applications.  
Connect a second cable from the other side of the SHV “Tee” to the high-power  
50 terminator. This cable should also be no longer than 1 m, although the  
requirement is less stringent than for the cable connecting the pulser output to the  
“Tee.”  
CW or Shutter Operation  
WARNING  
To prevent damage to the IIC-200 circuitry, the Berkeley Nucleonics Corp. Pulser  
should be disconnected for Shutter mode operation.  
For shuttering, set the Shutter/Gate switch to Shutter.  
Connect a BNC cable to the IIC-200 Shutter In BNC connector. Connect the other  
end to the Controller’s TTL SHUTTER OUT BNC connector (may also be named  
SHUTTER or SHUTTER MONITOR, depending on the controller model) .  
 
Version 1.A  
Model IIC-200 Operation Manual  
13  
Power Up Steps  
Make sure the front-panel MCP Power switch is Off. Adjust MCP Gain to minimum  
and turn on the chassis power.  
Note: As long as the image intensifier and IIC-200 are purchased from Roper  
Scientific as part of the same order, the upper limit circuit will be set for the  
maximum safe limit of the intensifier. In this case, the maximum voltage may be less  
than -1000 V.  
After the MCP voltage has been properly set, set the MCP Power switch on and  
slowly bring up the MCP voltage to the desired setting. High voltage will now be on  
the intensifier. It is normal for the alarm beeper to sound when power is first applied.  
However, if it continues to sound, set the MCP power switch off and determine the  
cause (e.g. switch set to Shutter instead of Gate or excessive light on the  
Photocathode).  
If the intensifier is operated at excessive MCP voltage, ion feedback may occur. If  
allowed to continue this condition may damage the intensifier. Therefore, be sure to  
monitor the light output from the intensifier (usually done by reading the CCD or other  
readout device) when increasing the MCP voltage setting. A sudden increase in output  
light, especially if not related to light on the input, is evidence of ion feedback. As  
intensifier tubes age, they sometimes become more susceptible to ion feedback, so it is  
advisable to check rather than to rely entirely on the factory limit setting.  
CAUTION  
For initial set-up, the Brightness Control feature can be used. Set the Auto Bright  
Control switch to the on position. This will reduce MCP voltage (and hence gain) in  
proportion to light level, so a greater range of input light can be tolerated. The alarm  
is still functional in this mode, and the intensifier voltage will still be shut down in  
case of excessive exposure levels. Be sure to set the Auto Bright Control switch off  
before taking data, or the intensifier will be nonlinear.  
The High Voltage Cable  
Two versions of the 6-pin high voltage cable have been shipped by Roper Scientific.  
Early high-voltage cables have a white housing. Later ones have a black housing and a  
white stripe. These cables are wired differently. The IIC-200 is compatible with the  
newer style cables only.  
If you have an older style Princeton Instruments image intensifier unit, originally  
supplied with the old-style cable, it will be necessary to return intensifier back to Roper  
Scientific to be updated before it can be used with the IIC-200. Intensifiers for use with  
black cables always have 6 conductors on the intensifier connector.  
WARNING  
The cable with the white housing is unsafe if not used strictly according to instructions.  
If a white cable is connected at one end only to an IIC-200 and the IIC-200 is turned on,  
arcing at the open end of the cable is likely and could cause painful and possibly  
hazardous electrical shock. Roper Scientific can update any white cable and 4 conductor  
image intensifier combination to the black-cable 6-conductor combination. Roper  
 
14  
Model IIC-200 Operation Manual  
Version 1.A  
Scientific highly recommends updating old units at the user’s earliest convenience.  
Contact the factory for details.  
Connecting the wrong cable at both ends can result in non-functioning of the image  
intensifier. If you suspect that you may be using the wrong cable for your intensifier,  
contact the factory or your local sales representative. If you have only purchased a single  
image intensifier and IIC-200 combination, the correct cable was shipped with your  
system.  
 
Warranty & Service  
Warranty  
This equipment is warranted to be free from defects of material and workmanship. It is  
sold subject to the mutual agreement that the liability of Roper Scientific is limited to  
replacing defective parts and/or repairing malfunctioning equipment at its factory,  
provided the equipment is returned, transportation prepaid, within twelve (12) months of  
its factory ship date.  
The purchaser agrees that Roper Scientific shall assume no liability for consequential  
damages resulting from its use or from packaging of shipments returned to the factory.  
Components which are damaged by misuse are not warranted. Units which have been  
modified by a customer are not warranted.  
UV coatings are not covered by this warranty.  
Equipment Repairs  
It is recommended that units requiring service in the United States be returned to the  
factory located in Trenton, New Jersey. Before instrumentation is returned for service,  
please consult a service engineer at the factory. In many cases, the problem may be  
cleared up over the telephone.  
If the unit needs to be returned, the service engineer will ask for a detailed explanation of  
the problems encountered and a purchase order to cover any charges. You will then  
receive a Returned Materials Authorization (RMA) number. Place this number on the  
package so the returned equipment can be easily identified when received at the factory.  
You must also include with the equipment a completed RMA form explaining the  
symptoms or problems encountered. Without this document, repair turnaround time will  
be considerably longer.  
If the unit is under warranty, the customer is only responsible for the transportation and  
insurance charges to Roper Scientific. Roper Scientific is responsible for the return  
transportation charges. If the unit is out of warranty, the customer is responsible for all  
transportation charges (including insurance and duty fees, when applicable) as well as all  
charges incurred to perform the repairs. In this case, the customer can decide the  
insurance value.  
International customers should contact your local manufacturers representative or  
distributor for repair information. See next page for contact information.  
15  
 
16  
Model IIC-200 Operation Manual  
Version 1.A  
Contact Information  
Roper Scientific’s manufacturing facility is located at the following address:  
Roper Scientific  
3660 Quakerbridge Road  
Trenton, NJ 08619 (USA)  
Tel: 609-587-9797  
Fax: 609-587-1970  
Tech Support E-mail: [email protected]  
For technical support and service outside the United States, see our web page at  
addresses of Roper Scientific’s overseas offices and representatives is maintained on the  
web page.  
 
Index  
Alarm circuit, 10  
Audio beeper, 7  
Intensifier H.V.P.S. cable  
versions, 13  
Auto Bright Control switch, 8  
Auto Brightness Control, 7, 10  
Beeper, 7  
Bias levels  
MCP-In, 7  
warnings, 13  
Intensifier H.V.P.S. input, 8  
Line voltage selection, 9  
MCP Gain dial, 8  
MCP power supply, 7  
MCP Power switch, 8  
Operation instructions, 11  
gated operation (Gen II), 11  
gated operation (Gen IV), 12  
Gen II, 11  
Gen IV, 12  
power up, 13  
Shutter mode operation (Gen II, 11  
Overcurrent condition  
beeper, 7  
MCP-Out, 7  
phosphor, 7  
Brightness control, 7, 10  
Contact information, 16  
e-mail address, 16  
street address, 16  
teelphone and fax numbers, 16  
web site, 16  
CW mode, 7  
CW operation (Gen IV), 12  
Fuses, 9  
shutdown, 7  
Gain reduction, 7  
Gate In input, 8  
Gate mode, 8  
Power (IIC-200) switch, 8  
Power module, 8, 9  
Power Up instructions, 13  
Repair policy, 15  
Gate pulse  
Gen II intensifier, 7  
Gen IV intensifier, 7  
Gated operation  
sample/hold, 10  
Shutter In input, 8  
Shutter mode, 8  
Gen II, 11  
Gen IV, 12  
Shutter mode operation  
Gen II, 11  
Gen II intensifier, 7  
connection instructions, 11  
gated operation, 11  
shutter mode operation, 11  
Gen IV intensifier, 7  
connection instructions, 12  
CW operation, 12  
gated operation, 12  
shutter mode operation, 12  
High voltage cable, 11  
versions, 13  
Shutter pulse, 7  
Shutter/Gate switch, 8  
Switches  
AutoBright Control, 8  
MCP Power switch, 8  
power (IIC-200), 8  
Shutter/Gate, 8  
Technical support  
e-mail address, 16  
telephone and fax numbers, 16  
web site address, 16  
Voltage  
warnings, 13  
High voltage input, 8  
High voltage power supply (H.V.P.S.), 8  
High voltage shutdown, 10  
Inputs  
photocathode to MCP, 7  
Voltages  
IIC-200 power modules, 8  
intensifier, 8  
Gate In, 8  
high voltage, 8  
Intensifier H.V.P.S., 8  
Shutter In, 8  
intensifier H.V.P.S., 8  
line voltage selection, 9  
Warranty, 15  
17  
 
18  
Model IIC-200 Operation Manual  
Version 1.A  
This page intentionally left blank.  
 

RCA Flat Panel Television D52W17 User Guide
RIDGID Saw R4516 User Guide
Rimage Printer 110660 001_F User Guide
RocketFish Speaker System RF 2990 User Guide
Samson Stereo Amplifier PB9 User Guide
Samsung All in One Printer SCX 4216F User Guide
Samsung Bread Maker SGH D780 User Guide
SanDisk Computer Drive 80 36 00323 User Guide
Sears Edger 53679751 User Guide
SEH Computertechnik Planer IC146 ETHER HP FL User Guide