The OMEGA PPA100 Hand  
Pump is designed to be a simple  
to use pressure and vacuum  
source for calibrating process  
control instrumentation.  
Combined with one of our  
pressure calibrators, the PPA100  
makes a complete Pressure/  
Vacuum Calibration System.  
Ope ration  
The drawing shows the  
operation of the controls on the  
PPA100. The pump is designed  
for use with ambient air only.  
4. Repeat the process several  
times using a new cotton  
swab each time until there is  
no remaining evidence of  
contamination or dirt.  
8. Once all parts have been  
cleaned and inspected,  
reinstall the o-ring and spring  
assembly into the valve body.  
9. Reinstall the retention caps  
and gently tighten each cap.  
Pump Valve As s e mbly  
Cle aning Ins tructions  
Occasionally, the PPA100 may  
not work properly due to dirt or  
other contamination of the  
internal valve assembly. Use the  
following procedure for cleaning  
the valve assembly.  
5. Operate the pump handles  
several times and recheck for 10. Seal the output port and  
contamination.  
6. Clean the o-ring assembly  
and the o-ring on the  
operate the pump to at least  
50% of capacity.  
11. Release the pressure and  
repeat several times to  
ensure that the o-rings seat  
properly.  
retention caps with IPA and  
inspect the o-rings closely for  
any damage or excessive  
wear. Replacements are  
included in the repair kit, if  
needed.  
1. Using a small screwdriver,  
remove the 2 valve retention  
caps located on opposite  
sides of the pump below the  
pressure/vacuum switch.  
2. After the caps have been  
removed, gently remove the  
spring and o-ring assembly.  
3. Set aside the valve  
7. Inspect the springs for wear  
or loss of tension. They  
should be approximately 8.6  
mm long in the relaxed state.  
If shorter, they may not  
provide sufficient sealing  
tension. Replace if needed.  
assemblies in a safe area  
and clean out the valve body  
using a cotton swab soaked  
in IPA (isopropyl alcohol).  
 

Miele Beverage Dispenser KWT4154UG1 User Manual
Nlynx Network Router MinIT6000 User Manual
Omega Engineering Water Pump FPUHP User Manual
Omnimount Indoor Furnishings OM10328 User Manual
One for All Universal Remote URC 4080 User Manual
Optoma Technology Projector EP783 User Manual
Oster Toaster TSSTRTS2S1 User Manual
Panasonic Digital Camera DMC F3 User Manual
Panasonic Fax Machine KX F220 User Manual
Panasonic Portable DVD Player CQVX100U User Manual